MDCG 2025-10: Guidance on Post-Market Surveillance of Medical Devices and IVDs
Document: MDCG 2025-10
Year: 2025
Published: 19 December 2025
Regulation: EU MDR 2017/745 / EU IVDR 2017/746
Official PDF: Download
Purpose
This guidance provides a comprehensive description of the Post-Market Surveillance (PMS) system required under the EU MDR (Article 83) and IVDR (Article 78). It clarifies expectations for PMS plans, main PMS activities, and the interactions between PMS and other key aspects of the Quality Management System (QMS).
Key Points
- PMS as a continuous, proactive system: PMS is not a reactive checklist but a continuous, structured system embedded within the manufacturer's QMS. Surveillance begins when the first device is placed on the market and continues until the end of the intended lifetime of the last device.
- Proactive data collection: Manufacturers must actively seek out information and not merely wait for complaints. This includes customer surveys, clinical experience, user feedback, literature screening, registries, and post-market studies.
- PMS Plan requirements: Each device or device group must be covered by a PMS plan that specifies methods, data sources, frequencies, threshold values, trend reporting methodology, and PMCF/PMPF plans.
- Indicators and threshold values: Suitable indicators and thresholds must be established pre-market and used for continuous reassessment of the benefit-risk ratio.
- Trend reporting: The PMS plan must outline statistical methods for detecting any statistically significant increase in frequency or severity of non-serious incidents, as well as observation periods and reporting procedures.
- QMS integration: PMS data must actively drive decisions and feed into risk management, clinical/performance evaluation, design and labelling updates, SS(C)P, CAPA, and management review.
- Custom-made devices: PMS requirements under Article 83 MDR apply to custom-made devices; PMCF plans are required.
- One plan for multiple devices: A single PMS plan can cover devices with the same manufacturing process, design, intended purpose, or device family.
- SOTA assessment: PMS data should be used to confirm the device remains consistent with the generally accepted State of the Art.
- Top management reporting: The effectiveness of the PMS system should be reported to top management, e.g., via management review.
Structure
| Section | Content |
|---|---|
| 1 | Introduction |
| 2 | Scope and Objectives |
| 3 | The PMS System required by MDR/IVDR (general obligations + custom-made devices) |
| 4 | The PMS plan (Table 1: elements required per Annex III) |
| 5 | Main activities of the PMS system (data sources, collection, assessment, conclusions) |
| 6 | Interactions of PMS with other QMS aspects (Table 3: risk management, clinical evaluation, SS(C)P, CAPA, design, trend reporting) |
| 7 | References |
| Annex 1 | Overview of PMS obligations in MDR and IVDR |
| Annex 2 | Scenarios: IVD and medical device examples of PMS data utilization |
Relationship to Other MDCG Documents
| Document | Relationship |
|---|---|
| MDCG 2022-21 | PSUR guidance -- MDCG 2025-10 does not cover PSUR/PMS report preparation; refers to MDCG 2022-21 |
| MDCG 2020-7 | PMCF Plan Template -- referenced for PMCF/PMPF plan requirements |
| MDCG 2023-1 | Health institution exemption (out of scope of MDCG 2025-10) |
| MDCG 2023-3 | Vigilance terms and concepts -- referenced for serious incident reporting |
| MDCG 2021-3 | Custom-made devices Q&A -- referenced for CMD PMS obligations |
| MDCG 2024-1 | Device Specific Vigilance Guidance (DSVG) Template |
Key Standards Referenced
- EN ISO 13485:2016 (Quality management systems)
- EN ISO 14971:2019 (Risk management)
- CEN ISO/TR 20416:2020 (Post-market surveillance for manufacturers)
Official text
MDCG 2025-10
Guidance on post-market surveillance of medical devices and in vitro diagnostic medical devices
December 2025
This document has been endorsed by the Medical Device Coordination Group (MDCG) established by Article 103 of Regulation (EU) 2017/745. The MDCG is composed of representatives of all Member States and it is chaired by a representative of the European Commission. The document is not a European Commission document and it cannot be regarded as reflecting the official position of the European Commission. Any views expressed in this document are not legally binding and only the Court of Justice of the European Union can give binding interpretations of Union law.
1. Introduction
Regulation (EU) 2017/745 on medical devices (MDR) and Regulation (EU) 2017/746 on in-vitro diagnostic medical devices (IVDR) establish a regulatory framework to ensure a high level of protection of the health and safety of patients, users and other individuals, whilst supporting innovation. Devices may only be placed on the Union market if they comply with these Regulations.
To ensure that devices remain compliant with the requirements of the Regulations and that real-world experience gained from the use of the devices is taken into account for the product realisation, manufacturers must have a Quality Management System (QMS) comprising a Post-Market Surveillance (PMS) system in place. These systems should be proportionate to the risk posed by the device and appropriate for the type of the device.
According to, respectively, Recitals 74 MDR and 75 IVDR, Article 10(10) of the MDR/IVDR, manufacturers are required to implement and keep up to date a PMS system in accordance with Article 83 MDR/Article 78 IVDR. This obligation reinforces the importance of a structured and proactive PMS approach as an integral part of the manufacturer's QMS.
Manufacturers need to play an active role during the post-market phase by:
- Systematically and actively gathering information from post-market experience with their devices.
- Using this information to update their technical documentation.
- Cooperating with national competent authorities responsible for vigilance and market surveillance.
To support this, manufacturers must:
- Establish a comprehensive PMS system, set up under their QMS.
- Base the PMS system on a PMS plan.
Data and insights from PMS activities must be used to:
- Update relevant parts of the technical documentation (e.g., risk assessment, clinical evaluation).
- Identify the need for preventive and/or corrective action.
- Enhance transparency (e.g. through the SS(C)P).
PMS is a continuous process. Manufacturers must carry out PMS throughout the lifetime of a device to monitor its safety, quality and performance.
After having established the initial PMS plan, which is part of the technical documentation, the first PMS process cycle begins when the first device has been placed on the market/put into service and concludes with a report. However, surveillance activities and data collection must be systematically performed and recorded on a continuous, ongoing basis, throughout the entire period between the first placing on the market/putting into service and the end of the intended lifetime of the last device placed on the market/put into service.
2. Scope and Objectives
Unless otherwise stated, this guidance is applicable to all medical devices (MDs) and in vitro diagnostic medical devices (IVDs). The main objectives of this guidance are:
- To describe the PMS system.
- To describe the PMS plan.
- To describe the main activities within the PMS system.
- To clarify the interactions of the PMS system in accordance with Article 83 MDR/Article 78 IVDR with other key aspects of the QMS as described in Article 10(9) MDR and Article 10(8) IVDR.
Out of scope:
- This guidance does not provide details on how a manufacturer should prepare a periodic safety update report ('PSUR') or a post-market surveillance report. Specific guidance on PSUR is provided in MDCG 2022-21. Although not covering post-market surveillance reports, MDCG 2022-21 may provide useful suggestions on how a manufacturer can present information in a post-market surveillance report.
- This guidance does not cover the requirements for health institution exemption under Article 5(5) MDR/IVDR (in-house devices), though it is expected that health institutions review experience gained from the use of in-house devices and take all necessary corrective actions.
3. The PMS System required by the MDR/IVDR
Article 10(9)(i) MDR and Article 10(8)(i) IVDR refer to a PMS system in accordance with Article 83 MDR/Article 78 IVDR. Manufacturers are required to 'plan, establish, document, implement, maintain and update' a PMS system which is to be implemented effectively and in a manner 'that is proportionate to the risk class and appropriate for the type of device'.
3.1 General obligations
The PMS system must be suited to actively and systematically gathering, recording and analysing relevant data on the quality, performance and safety of a device throughout its entire lifetime. It must also enable the manufacturer to draw the necessary conclusions and support the determination, implementation and monitoring of any preventive or corrective actions.
In addition to this requirement, it is in the manufacturer's interest to use the PMS data to identify options to improve the usability, performance, and safety of the device.
The PMS system interfaces directly with other aspects of the QMS, such as clinical/performance evaluation and risk management.
For a successful implementation of the PMS system the manufacturer must provide for appropriate structures, procedures, processes and resources. The effectiveness of the PMS system should be reported to top management, e.g., via management review.
3.2 Specific obligations for manufacturers of custom-made devices under the MDR
The PMS requirements according to Article 83 MDR are applicable to all devices including Custom-Made Devices (CMD). Section 5 of Annex XIII MDR states that "The manufacturer shall review, and document experience gained in the post-production phase, including from PMCF as referred to in Part B of Annex XIV MDR, and implement appropriate means to apply any necessary corrective action."
Post-market review is mainly focused on the expected performance. A PMCF plan is therefore required. This implies that for a CMD, the manufacturer needs to establish and maintain a PMS system, and to document the experience gained from the post-production phase.
For class I devices, the CMD manufacturer must establish a PMS report according to Article 85 MDR whereas, for classes IIa, IIb and III devices, a Periodic Safety Update Report (PSUR) according to Article 86 MDR must be established.
To implement a MDR compliant post-market surveillance system, the CMD manufacturer should establish appropriate communication channels with relevant healthcare providers/professionals or patients to receive feedback on the quality, performance and in particular the clinical performance and safety of the devices in the field.
For risk management, post-market surveillance and clinical evaluation life cycle processes as defined by the MDR, CMD manufacturers should apply these obligations to groups of devices with the same intended purpose, materials and processes used, principal design etc. and not to each individual CMD.
4. The PMS plan
Article 84 MDR/Article 79 IVDR refer to the PMS system being based on a PMS plan, which is part of the technical documentation (except for custom-made devices under the MDR). The requirements for the PMS plan are outlined in section 1 of Annex III of both the MDR and IVDR.
PMS planning starts already during the development of a device. The manufacturer should think ahead and determine which activities should be carried out to systematically and proactively collect experience gained from devices they place on the market or put into service.
Each device must be covered by a PMS plan. A plan can cover a single device or a group of devices. For example, devices with the same manufacturing process, design and intended purpose, or in the same device family can be covered by the same PMS plan. The plan should clearly state which devices are included within its scope.
The PMS plan specifies the aspects of the device or group of devices to be monitored, the frequency of monitoring, and the methods to be applied. The chosen methods should be appropriate and based on the device's risk profile, and the rationale for selecting these methods needs to be documented.
Although the fourth indent of section 1(b) of Annex III MDR/IVDR requires that methods and procedures are covered in the PMS plan, in practical terms, the PMS plan should define what types of methods are to be applied. An explanation of how and by whom these methods are to be applied may instead be covered by procedures that are referenced within the PMS plan.
The concept of "proactive" is central to understand the requirements set out in section 1(b) of Annex III MDR/IVDR. "Proactive" emphasizes the manufacturer's obligation to actively seek out available information and not merely wait for it to arrive through channels such as complaints. Manufacturers should play an active role during the post-market phase by systematically and actively gathering information from post-market experience with their devices.
This involves deliberately planning the gathering or generating information from a variety of sources. These may include, but are not limited to, customer surveys, gathering of clinical experience, user feedback, screening of scientific literature, other clinical data sources, meta-analyses of published clinical data, evaluation of suitable registers, and post-market studies.
The PMS plan should also reference procedures and frequencies related to the output of the plan, such as a PMS report for class I devices per Article 85 MDR and a PSUR for all other classifications per Article 86 MDR/Article 81 IVDR.
Table 1: Summary of the elements to be covered in the PMS plan per section 1(b) of Annex III MDR/IVDR
The PMS plan must include:
| Element | Explanation | Examples |
|---|---|---|
| Proactive and systematic process to collect information | Manufacturers are expected to collect information to address understanding of device usage, safety and performance, usability, and identify areas for improvement. The information is used to demonstrate that the device remains state of the art (SOTA). | List of activities with references to procedures; sources of data regarding similar products; frequency and responsible roles |
| Effective and appropriate methods and processes to assess the collected data | Methods and processes for assessing the data are expected to be proportionate to the type of device. Higher-risk devices will require more specific methods of data analysis. | Methods to be applied (qualitative, quantitative, statistical); list of parameters to be analysed; list of measurable values |
| Suitable indicators and threshold values | Suitable indicators and threshold values (limits beyond which action should be taken) are to be established in the pre-market phase. Data obtained through PMS feeds back into risk management. | Indicators and thresholds relevant for assessing impact on hazards, frequency of occurrence, associated risks, benefit-risk ratio |
| Effective methods and tools to investigate complaints and analyze market-related experience | Methods and tools proportionate to the type of device. Higher-risk devices require more rigorous approaches. | Methods including those described in IMDRF Adverse Event Terminology Annex B |
| Methods and protocols to manage incidents subject to trend report (Article 88 MDR/Article 83 IVDR) | Methodology for determining any statistically significant increase in frequency or severity, as well as observation period. | References to procedures and statistical methods; threshold values and indicators; frequency and responsible roles |
| Methods and protocols to communicate effectively with competent authorities, notified bodies, economic operators, and users | Processes to ensure effective communication with respect to the output of PMS activities. | References to procedures; communication methods and data transfer tools |
| Reference to procedures to fulfil manufacturer obligations laid down in Articles 83, 84 and 86 MDR | Procedures for implementation of PMS system, and for generating PMS-related plans and reports. | References to relevant procedures; frequency and responsible roles |
| Systematic procedures to identify and initiate appropriate measures including corrective actions | Procedures describing criteria for corrective actions and the type of measure to be undertaken. | List of identified criteria; responsible roles |
| Effective tools to trace and identify devices for which corrective actions might be necessary | Tools for identifying and tracing devices. Procedures for non-conforming devices (recall and quarantine). | Methods/tools to trace and identify devices; references to procedures |
| PMCF/PMPF plan, or justification as to why not applicable | The PMCF/PMPF plan is an integral part of the PMS plan. If not required, a justification is needed. | Description of planned activities; reference to detailed plan/protocol; justification if not applicable |
5. The main activities of the PMS system
The PMS system should be suited to actively and systematically gathering, recording, and analysing relevant data on the quality, performance, and safety of a device throughout its entire lifetime.
5.1 Determining appropriate sources of available information for PMS
The manufacturer should decide which sources are relevant and appropriate for the device concerned. When selecting data sources, the different economic operators, users (healthcare professionals, patient, or lay users) and the situation in which the devices are used need to be considered.
Data quality and integrity should be considered before analysing data to ensure the information is reliable. For example, the use of unverifiable data can lead to over-reaction, as it can be based on non-scientific data sources such as social and public media.
Table 2: Information to be collected and use of that information (Section 1(a) Annex III MDR/IVDR)
| Required Information | Examples of information to be collected | Use of information |
|---|---|---|
| Information concerning serious incidents, including information from PSURs, and field safety corrective actions | Outputs from procedures to receive, document, investigate and analyze serious incidents; information/feedback from FSCAs | Identification of emerging risks; demonstration of continued acceptability; tracking SOTA developments; determination of reportability; identification of trends; comparison with similar products; identification of new benefits; identification of misuse/off-label use |
| Records referring to non-serious incidents and data on any undesirable side-effects | Incidents determined to be non-serious; complaints and information; service and repair records | Same as above, plus input to CAPA, FSCA, design changes, risk management updates |
| Information from trend reporting | Outputs from trending of non-serious incidents, undesirable side effects | Trend reporting per Article 88 MDR/Article 83 IVDR |
| Relevant specialist or technical literature, databases and/or registers | Literature review/clinical evaluation process; searching competent authority websites and EUDAMED; device registers or registries; clinical guidelines, HTA reports | Update of technical documentation, clinical/performance evaluation, SS(C)P, PMS plan |
| Information including feedback and complaints provided by users, distributors, and importers | Customer surveys; feedback from training/workshops; complaints other than serious incidents; feedback from sales channels | Preparation of PMS report/PSUR, PMCF/PMPF evaluation report, clinical/performance evaluation report |
| Publicly available information about similar devices | Literature screening; competent authority databases; device registers; IFU and labelling; SS(C)P; information from third countries; conferences | Comparison and SOTA assessment |
5.2 Collecting necessary data
The data collection for PMS typically starts once a manufacturer declares the device's conformity and places it on the market. When gathering PMS data, manufacturers should establish the processes for receiving, reviewing, and evaluating information, including the screening of scientific literature, feedback and complaints provided by users, distributors, importers or other third parties.
In some instances, manufacturers may be required to generate data relating to real-world use of the device. This is especially applicable in the case of higher risk devices and/or devices with novel features or applications, in most cases through specific methods and procedures of PMCF/PMPF.
5.3 Assessment and analysis of data
The PMS system should include the manufacturer's processes and methods that are used to effectively assess and analyse the data collected under the PMS plan. Data is assessed to analyse and characterise the performance of the device and to enable comparisons with similar devices, thus allowing the manufacturer to confirm the continued acceptability of the benefit-risk ratio.
The manufacturer should also reflect on the performance of their device versus the generally accepted State of the Art (SOTA), taking into consideration datasets relating to different patient populations, device combinations and/or models and variants.
Furthermore, when assessing the data, it is important that data from different sources is compared, and conflicting data identified and evaluated. Data quality and validity should be evaluated to enable the drawing of appropriate conclusions and where necessary to identify suitable actions.
5.4 Drawing conclusions and determining the need to take action(s)
Drawing the necessary conclusions and determining the need to take any preventive and corrective actions are central aspects of the manufacturer's assessment of the overall device safety and performance.
The drawing of conclusions should be based on the analysis of the data collected. Those conclusions and the actions taken subsequently need to be documented in (an update of) a summary report, either in the PMS report or in the PSUR.
Following the outcome of that report, a review of the current PMS plan may be necessary. This will drive the level of PMS activity required for the next cycle of PMS.
6. The interactions of the PMS system with other key aspects of the QMS
Article 83(3) MDR/Article 78(3) IVDR specifically request that information generated from PMS activities should be used as input, on a continuous basis, to other processes within the QMS.
Table 3: Interactions of the PMS system with other key aspects of the QMS
| QMS Process | Explanation | Activity |
|---|---|---|
| Update the benefit-risk determination and the risk management (Art. 83(3)(a) MDR/Art. 78(3)(a) IVDR) | Information obtained through PMS activities should be used in the benefit-risk determination process and in the continuous assessment of the overall benefit-risk profile. Suitable indicators and threshold values used in the continuous reassessment should be covered in the PMS plan and linked to risk management documentation. | Establish procedures for PMS-risk management interface; analyse and appraise PMS data; assess impact on probability and severity ratings; evaluate impact on overall risk and benefit-risk ratio |
| Update the design and manufacturing information, IFU and labelling (Art. 83(3)(b) MDR/Art. 78(3)(b) IVDR) | Information obtained through PMS should be used to evaluate the need for updating design and manufacturing information, IFU and labelling. Design and development process should consider PMS data as input. | Take appropriate action when evidence of new or increased risks is identified, including updates to device design, labelling and IFU |
| Update clinical evaluation / performance evaluation (Art. 61 and Art. 83(3)(c) MDR/ Art. 56 and Art. 78(3)(c) IVDR) | PMCF/PMPF should be addressed in the PMS plan and performed pursuant to a documented method. Under MDR, clinically relevant information from PMS (particularly PMCF data) is considered "clinical data". PMCF must be understood as a continuous process that updates the clinical evaluation. | Consider PMS outputs in clinical/performance evaluation updates; analyse PMCF/PMPF findings; document results in evaluation reports; consider data from overall benefit-risk perspective |
| Update the summary of safety and (clinical) performance (SS(C)P) (Art. 32 and Art. 83(3)(d) MDR/Art. 29 and Art. 78(3)(d) IVDR) | Applicable to class III and implantable devices (MDR) and class C and D (IVDR). The SS(C)P provides transparency of clinical data to healthcare professionals and patients. | Describe how PMS data updates the SS(C)P; assess impact on benefit-risk profile; align with CER/PER and PSUR |
| Identify needs for preventive, corrective, or field safety corrective actions (Art. 83(3)(e) MDR/Art. 78(3)(e) IVDR) | Manufacturers should set up PMS to actively collect, analyse and utilise relevant data. Draw conclusions as to whether corrective or preventive actions are required. | Integrate PMS data into QMS; initiate and document CAPA and FSCA |
| Identify options to improve usability, performance and safety (Art. 83(3)(f) MDR/Art. 78(3)(f) IVDR) | PMS information should be used to identify improvement options through linkage to design and development. These actions are proactive and strategic, not necessarily tied to identified risks. | Ensure design and development process defines PMS as input |
| Contribute to PMS of other devices (Art. 83(3)(g) MDR/Art. 78(3)(g) IVDR) | PMS data regarding one device can contain useful information for other devices sharing similar purpose, design or characteristics. | Use collected information as input to PMS of other devices |
| Detect and report trends (Art. 83(3)(h) MDR and Art. 88 MDR/Art. 78(3)(h) and Art. 83 IVDR) | Information on non-serious incidents and expected undesirable side-effects should be analysed for any statistically significant increase in frequency or severity. | Review statistically significant increases that could impact benefit-risk profile; submit trend reports as required |
7. References
This guidance has been developed taking into consideration the following documents:
- EN ISO 13485:2016 Medical devices -- Quality management systems -- Requirements for regulatory purposes
- EN ISO 14971:2019 Medical devices -- Application of risk management to medical devices
- EN ISO 14155:2020 Clinical investigation of medical devices for human subjects - Good clinical practice
- EN ISO 20916:2024 In vitro diagnostic medical devices -- Clinical performance studies using specimens from human subjects -- Good study practice
- EN 62366-1:2015 Medical devices -- Part 1: Application of usability engineering to medical devices
- CEN ISO/TR 20416:2020 Medical devices -- Post-market surveillance for manufacturers
- Commission Implementing Decision (EU) 2021/1182
- Guidance for post-market surveillance and market surveillance of medical devices, including in vitro diagnostics (WHO)
- IMDRF/GRRP WG/N47 Essential Principles of Safety and Performance of Medical Devices and IVD Medical Devices
Annex 1: Overview of PMS obligations in MDR and IVDR
| Element | Description | Obligations |
|---|---|---|
| PMS System (Art. 83 MDR/Art. 78 IVDR) | Comprehensive system to collect experience from use of devices. Applicable for all device classes. | Proactive and systematic approach; analyse data throughout lifetime; draw conclusions; connect with CAPA, vigilance, design, labelling; update technical documentation; integral part of QMS |
| PMS Plan (Art. 84 & Annex III MDR/Art. 79 & Annex III IVDR) | Describes implementation of PMS system. Applicable for all device classes. | Part of QMS and technical documentation; defines data to be collected; describes methods for monitoring trends, investigating complaints, reassessing benefit-risk, tracing devices; includes PMCF/PMPF plan or justification |
| PMS Report (Art. 85 MDR/Art. 80 IVDR) | Summary of results and conclusions. MDR: class I devices; IVDR: class A and B devices. | Summary of results and conclusions; rationale and description of any CAPA; update according to PMS plan; available to competent authority upon request |
| PSUR (Art. 86 MDR/Art. 81 IVDR) | Summary of results and conclusions. MDR: class IIa, IIb and III; IVDR: class C and D. | Detailed guidance in MDCG 2022-21 |
Annex 2: Scenarios -- how to use PMS data to update other processes
IVD Example: Increase in serious incidents due to incorrect values provided by blood glucose test strips. Workflow: Review and analyse cases -> Investigation shows out-of-specification test strips -> CAPA and FSCA required -> Recall affected test strips -> Submit FSCA report to CA -> Improve manufacturing process -> Assess impact to other IVDs.
Medical Device Example: Increase in complaints related to premature battery depletion in implantable cardioverter defibrillators (ICDs). Workflow: Review and analyse complaints -> Investigation shows lithium deposition causing short circuits -> New risk identified (abrupt stoppage leading to death) -> Update Risk Management File -> CAPA and FSCA required -> Recall affected devices -> Inform supplier and initiate design change -> Update design/manufacturing information, IFU and labelling -> Update clinical evaluation and SSCP -> Assess impact to other devices -> Consider trend reporting.

